Abstract
The characteristics of a computer-controlled Josephson voltage standard developed for routine calibrations of electrochemical cells are described. The device is also suitable for use as a transfer standard. The overall relative uncertainty (1 sigma ) of the device is less than one part of 10/sup 8/.
Original language | English |
---|---|
Pages (from-to) | 2-9 |
Journal | IEEE Transactions on Instrumentation and Measurement |
Volume | 37 |
Issue number | 1 |
DOIs | |
Publication status | Published - 1988 |
MoE publication type | A1 Journal article-refereed |