Abstract
An AC voltage standard is being developed based on phase sensitive detection of the amplitude of the fundamental frequency component of the output of a programmable Josephson voltage array. The setup is described and requirements for relative uncertainties less than 10/sup -7/ at 1 kHz and 1 V are discussed. According to preliminary experiments, the constructed current bias is able to drive the array from -1 to +1 V within less than 100 ns.
| Original language | English |
|---|---|
| Pages (from-to) | 533-537 |
| Number of pages | 5 |
| Journal | IEEE Transactions on Instrumentation and Measurement |
| Volume | 52 |
| Issue number | 2 |
| DOIs | |
| Publication status | Published - 2003 |
| MoE publication type | A1 Journal article-refereed |
Keywords
- AC voltage
- programmable Josephson voltage standard