An absolute pressure balance with a removable weight exchange system

Heikki Kajastie (Corresponding Author), M. Rantanen, K. Riski, S. Semenoja

    Research output: Contribution to journalArticleScientificpeer-review

    1 Citation (Scopus)

    Abstract

    The pressure balance used for MIKES barometer calibrations is a DH Instruments PG7601 for the range 5 kPa–350 kPa. The commonly used pressure range in barometer calibrations 850 hPa–1050 hPa has been automated by a removable weight exchange system. The new weight exchanger system has been tested in routine calibrations. The construction of the weight exchanger system for the pressure balance is described and the measurement results are shown.
    Original languageEnglish
    Pages (from-to)N33-N37
    Number of pages5
    JournalMeasurement Science and Technology
    Volume16
    Issue number10
    DOIs
    Publication statusPublished - 2005
    MoE publication typeA1 Journal article-refereed

    Fingerprint

    Dive into the research topics of 'An absolute pressure balance with a removable weight exchange system'. Together they form a unique fingerprint.

    Cite this