Abstract
The pressure balance used for MIKES barometer calibrations is a DH
Instruments PG7601 for the range 5 kPa–350 kPa. The commonly used
pressure range in barometer calibrations 850 hPa–1050 hPa has been
automated by a removable weight exchange system. The new weight
exchanger system has been tested in routine calibrations. The
construction of the weight exchanger system for the pressure balance is
described and the measurement results are shown.
| Original language | English |
|---|---|
| Pages (from-to) | N33-N37 |
| Journal | Measurement Science and Technology |
| Volume | 16 |
| Issue number | 10 |
| DOIs | |
| Publication status | Published - 2005 |
| MoE publication type | A1 Journal article-refereed |