Balance comparison for finnish accredited laboratories: 600 g and 10 kg balances

Maija Ojanen-Saloranta, J. Rintala, A. Paajanen, A. Rehula, M. Ijäs, L. Lindgren

    Research output: Chapter in Book/Report/Conference proceedingConference article in proceedingsScientificpeer-review

    Abstract

    VTT MIKES arranged a balance calibration comparison for Finnish laboratories accredited for balance calibration. The calibrated balances were 600 g ja 10 kg balances. Four accredited laboratories and seven persons carrying out the measurements participated in the comparison. All the results were in agreement with the reference value of the comparison. The largest normalized error En was 0.64.

    Original languageEnglish
    Title of host publication23rd Conference on the Measurement of Force, Mass and Torque 2017
    Subtitle of host publicationMeasurement Facing New Challenges - Held Together with TC5 and TC22
    PublisherCurran Associates Inc.
    Pages325-328
    ISBN (Electronic)978-1-5108-4471-1
    Publication statusPublished - 1 Jan 2017
    MoE publication typeA4 Article in a conference publication
    Event23rd IMEKO TC3 Conference on Measurement of Force, Mass and Torque 2017: Measurement Facing New Challenges - Helsinki, Finland
    Duration: 30 May 20171 Jun 2017

    Conference

    Conference23rd IMEKO TC3 Conference on Measurement of Force, Mass and Torque 2017: Measurement Facing New Challenges
    Country/TerritoryFinland
    CityHelsinki
    Period30/05/171/06/17

    Keywords

    • Accredited laboratory
    • Balance
    • Comparison

    Fingerprint

    Dive into the research topics of 'Balance comparison for finnish accredited laboratories: 600 g and 10 kg balances'. Together they form a unique fingerprint.

    Cite this