    Abstract

    VTT MIKES arranged a balance calibration comparison for Finnish laboratories accredited for balance calibration. The calibrated balances were 600 g ja 10 kg balances. Four accredited laboratories and seven persons carrying out the measurements participated in the comparison. All the results were in agreement with the reference value of the comparison. The largest normalized error En was 0.64.

    Original languageEnglish
    Title of host publication23rd Conference on the Measurement of Force, Mass and Torque, Together with 4th TC22 Conference on Vibration Measurement and 13th TC5 Conference on the Measurement of Hardness
    ISBN (Electronic)9781510844711
    Publication statusPublished - 1 Jan 2017
    MoE publication typeA4 Article in a conference publication
    Event23rd Conference on the Measurement of Force, Mass and Torque, Together with 4th TC22 Conference on Vibration Measurement and 13th TC5 Conference on the Measurement of Hardness - Helsinki, Finland
    Duration: 30 May 20171 Jun 2017

    Conference

    Conference23rd Conference on the Measurement of Force, Mass and Torque, Together with 4th TC22 Conference on Vibration Measurement and 13th TC5 Conference on the Measurement of Hardness
    CountryFinland
    CityHelsinki
    Period30/05/171/06/17

    Fingerprint

    Calibration

    Keywords

    • Accredited laboratory
    • Balance
    • Comparison

    Cite this

    Ojanen-Saloranta, M., Rintala, J., Paajanen, A., Rehula, A., Ijäs, M., & Lindgren, L. (2017). Balance comparison for finnish accredited laboratories: 600 g and 10 kg balances. In 23rd Conference on the Measurement of Force, Mass and Torque, Together with 4th TC22 Conference on Vibration Measurement and 13th TC5 Conference on the Measurement of Hardness
    Ojanen-Saloranta, M. ; Rintala, J. ; Paajanen, A. ; Rehula, A. ; Ijäs, M. ; Lindgren, L. / Balance comparison for finnish accredited laboratories : 600 g and 10 kg balances. 23rd Conference on the Measurement of Force, Mass and Torque, Together with 4th TC22 Conference on Vibration Measurement and 13th TC5 Conference on the Measurement of Hardness. 2017.
    @inproceedings{217b9ef074b44ce090d02badbc639248,
    title = "Balance comparison for finnish accredited laboratories: 600 g and 10 kg balances",
    abstract = "VTT MIKES arranged a balance calibration comparison for Finnish laboratories accredited for balance calibration. The calibrated balances were 600 g ja 10 kg balances. Four accredited laboratories and seven persons carrying out the measurements participated in the comparison. All the results were in agreement with the reference value of the comparison. The largest normalized error En was 0.64.",
    keywords = "Accredited laboratory, Balance, Comparison",
    author = "M. Ojanen-Saloranta and J. Rintala and A. Paajanen and A. Rehula and M. Ij{\"a}s and L. Lindgren",
    year = "2017",
    month = "1",
    day = "1",
    language = "English",
    booktitle = "23rd Conference on the Measurement of Force, Mass and Torque, Together with 4th TC22 Conference on Vibration Measurement and 13th TC5 Conference on the Measurement of Hardness",

    }

    Ojanen-Saloranta, M, Rintala, J, Paajanen, A, Rehula, A, Ijäs, M & Lindgren, L 2017, Balance comparison for finnish accredited laboratories: 600 g and 10 kg balances. in 23rd Conference on the Measurement of Force, Mass and Torque, Together with 4th TC22 Conference on Vibration Measurement and 13th TC5 Conference on the Measurement of Hardness. 23rd Conference on the Measurement of Force, Mass and Torque, Together with 4th TC22 Conference on Vibration Measurement and 13th TC5 Conference on the Measurement of Hardness, Helsinki, Finland, 30/05/17.

    Balance comparison for finnish accredited laboratories : 600 g and 10 kg balances. / Ojanen-Saloranta, M.; Rintala, J.; Paajanen, A.; Rehula, A.; Ijäs, M.; Lindgren, L.

    23rd Conference on the Measurement of Force, Mass and Torque, Together with 4th TC22 Conference on Vibration Measurement and 13th TC5 Conference on the Measurement of Hardness. 2017.

    Research output: Chapter in Book/Report/Conference proceedingConference article in proceedingsScientificpeer-review

    TY - GEN

    T1 - Balance comparison for finnish accredited laboratories

    T2 - 600 g and 10 kg balances

    AU - Ojanen-Saloranta, M.

    AU - Rintala, J.

    AU - Paajanen, A.

    AU - Rehula, A.

    AU - Ijäs, M.

    AU - Lindgren, L.

    PY - 2017/1/1

    Y1 - 2017/1/1

    N2 - VTT MIKES arranged a balance calibration comparison for Finnish laboratories accredited for balance calibration. The calibrated balances were 600 g ja 10 kg balances. Four accredited laboratories and seven persons carrying out the measurements participated in the comparison. All the results were in agreement with the reference value of the comparison. The largest normalized error En was 0.64.

    AB - VTT MIKES arranged a balance calibration comparison for Finnish laboratories accredited for balance calibration. The calibrated balances were 600 g ja 10 kg balances. Four accredited laboratories and seven persons carrying out the measurements participated in the comparison. All the results were in agreement with the reference value of the comparison. The largest normalized error En was 0.64.

    KW - Accredited laboratory

    KW - Balance

    KW - Comparison

    UR - http://www.scopus.com/inward/record.url?scp=85041462763&partnerID=8YFLogxK

    M3 - Conference article in proceedings

    AN - SCOPUS:85041462763

    BT - 23rd Conference on the Measurement of Force, Mass and Torque, Together with 4th TC22 Conference on Vibration Measurement and 13th TC5 Conference on the Measurement of Hardness

    ER -

    Ojanen-Saloranta M, Rintala J, Paajanen A, Rehula A, Ijäs M, Lindgren L. Balance comparison for finnish accredited laboratories: 600 g and 10 kg balances. In 23rd Conference on the Measurement of Force, Mass and Torque, Together with 4th TC22 Conference on Vibration Measurement and 13th TC5 Conference on the Measurement of Hardness. 2017