Abstract
VTT MIKES arranged a balance calibration comparison for Finnish laboratories accredited for balance calibration. The calibrated balances were 600 g ja 10 kg balances. Four accredited laboratories and seven persons carrying out the measurements participated in the comparison. All the results were in agreement with the reference value of the comparison. The largest normalized error En was 0.64.
Original language | English |
---|---|
Title of host publication | 23rd Conference on the Measurement of Force, Mass and Torque 2017 |
Subtitle of host publication | Measurement Facing New Challenges - Held Together with TC5 and TC22 |
Publisher | Curran Associates Inc. |
Pages | 325-328 |
ISBN (Electronic) | 978-1-5108-4471-1 |
Publication status | Published - 1 Jan 2017 |
MoE publication type | A4 Article in a conference publication |
Event | 23rd IMEKO TC3 Conference on Measurement of Force, Mass and Torque 2017: Measurement Facing New Challenges - Helsinki, Finland Duration: 30 May 2017 → 1 Jun 2017 |
Conference
Conference | 23rd IMEKO TC3 Conference on Measurement of Force, Mass and Torque 2017: Measurement Facing New Challenges |
---|---|
Country/Territory | Finland |
City | Helsinki |
Period | 30/05/17 → 1/06/17 |
Keywords
- Accredited laboratory
- Balance
- Comparison