Calculable impulse voltage calibrator for calibration of impulse digitizers

J. Hällström, Y. Chekurov, Martti M. Aro

Research output: Contribution to journalArticle in a proceedings journalScientificpeer-review

2 Citations (Scopus)

Abstract

Design and construction of a calibrator for lightning and switching impulse voltage measuring instruments is presented. The operating voltage range is from 50 mV to 300 V. The estimated uncertainty for the peak value is 0,03% and for front time and time to half value less than 1%.

Original languageEnglish
Pages (from-to)286-287
Number of pages2
JournalCPEM Digest (Conference on Precision Electromagnetic Measurements)
Publication statusPublished - 1 Jan 2002
MoE publication typeA4 Article in a conference publication
Event2002 Conference on Precision Electromagnetic Measurements - Ottawa, Ont., Canada
Duration: 16 Jun 200221 Jun 2002

Fingerprint

analog to digital converters
impulses
Calibration
lightning
Electric potential
electric potential
Lightning
Uncertainty

Cite this

@article{c9a2e73b2658409c9d92e8f6e6a9a0c5,
title = "Calculable impulse voltage calibrator for calibration of impulse digitizers",
abstract = "Design and construction of a calibrator for lightning and switching impulse voltage measuring instruments is presented. The operating voltage range is from 50 mV to 300 V. The estimated uncertainty for the peak value is 0,03{\%} and for front time and time to half value less than 1{\%}.",
author = "J. H{\"a}llstr{\"o}m and Y. Chekurov and Aro, {Martti M.}",
year = "2002",
month = "1",
day = "1",
language = "English",
pages = "286--287",
journal = "CPEM Digest (Conference on Precision Electromagnetic Measurements)",
issn = "0589-1485",
publisher = "IEEE Institute of Electrical and Electronic Engineers",

}

Calculable impulse voltage calibrator for calibration of impulse digitizers. / Hällström, J.; Chekurov, Y.; Aro, Martti M.

In: CPEM Digest (Conference on Precision Electromagnetic Measurements), 01.01.2002, p. 286-287.

Research output: Contribution to journalArticle in a proceedings journalScientificpeer-review

TY - JOUR

T1 - Calculable impulse voltage calibrator for calibration of impulse digitizers

AU - Hällström, J.

AU - Chekurov, Y.

AU - Aro, Martti M.

PY - 2002/1/1

Y1 - 2002/1/1

N2 - Design and construction of a calibrator for lightning and switching impulse voltage measuring instruments is presented. The operating voltage range is from 50 mV to 300 V. The estimated uncertainty for the peak value is 0,03% and for front time and time to half value less than 1%.

AB - Design and construction of a calibrator for lightning and switching impulse voltage measuring instruments is presented. The operating voltage range is from 50 mV to 300 V. The estimated uncertainty for the peak value is 0,03% and for front time and time to half value less than 1%.

UR - http://www.scopus.com/inward/record.url?scp=0036388618&partnerID=8YFLogxK

M3 - Article in a proceedings journal

AN - SCOPUS:0036388618

SP - 286

EP - 287

JO - CPEM Digest (Conference on Precision Electromagnetic Measurements)

JF - CPEM Digest (Conference on Precision Electromagnetic Measurements)

SN - 0589-1485

ER -