Calculable impulse voltage calibrator for calibration of impulse digitizers

J. Hällström, Y. Chekurov, Martti M. Aro

Research output: Contribution to journalArticle in a proceedings journalScientificpeer-review

4 Citations (Scopus)

Abstract

Design and construction of a calibrator for lightning and switching impulse voltage measuring instruments is presented. The operating voltage range is from 50 mV to 300 V. The estimated uncertainty for the peak value is 0,03% and for front time and time to half value less than 1%.

Original languageEnglish
Pages (from-to)286-287
Number of pages2
JournalCPEM Digest (Conference on Precision Electromagnetic Measurements)
Publication statusPublished - 1 Jan 2002
MoE publication typeA4 Article in a conference publication
EventConference on Precision Electromagnetic Measurements, CPEM 2002 - Ottawa, Canada
Duration: 16 Jun 200221 Jun 2002

Fingerprint

Dive into the research topics of 'Calculable impulse voltage calibrator for calibration of impulse digitizers'. Together they form a unique fingerprint.

Cite this