Calibration of a Two-Port Millimeter-wave Quasioptical Measurement System

Maxim Masyukov*, Aleksi Tamminen, Irina Nefedova, Juha Ala-Laurinaho, Samu Ville Palli, Andrey Generalov, Zachary Taylor

*Corresponding author for this work

Research output: Chapter in Book/Report/Conference proceedingConference abstract in proceedingsScientific

Abstract

A quasioptical setup based on a Gaussian beam telescope system has been created to analyze the room-temperature and cryogenic millimeter-wave S-parameters of materials and devices for astronomical instrumentation, simulated and tested. The room temperature tests were performed in the WR-3.4 (220-330 GHz) frequency range after completing the thru-reflect-line calibration with a set of custom calibration standards. The system capabilities have been tested during the characterization of the planar devices and materials.

Original languageEnglish
Title of host publication2024 49th International Conference on Infrared, Millimeter, and Terahertz Waves, IRMMW-THz 2024
PublisherIEEE Institute of Electrical and Electronic Engineers
ISBN (Electronic)979-8-3503-7032-4
ISBN (Print)979-8-3503-7033-1
DOIs
Publication statusPublished - 7 Oct 2024
MoE publication typeNot Eligible
Event49th International Conference on Infrared, Millimeter, and Terahertz Waves, IRMMW-THz 2024 - Perth, Australia
Duration: 1 Sept 20246 Sept 2024

Conference

Conference49th International Conference on Infrared, Millimeter, and Terahertz Waves, IRMMW-THz 2024
Country/TerritoryAustralia
CityPerth
Period1/09/246/09/24

Keywords

  • Calibration
  • cryogenics
  • instrumentation
  • millimeter-waves
  • quasioptics

Fingerprint

Dive into the research topics of 'Calibration of a Two-Port Millimeter-wave Quasioptical Measurement System'. Together they form a unique fingerprint.

Cite this