Calibration of dial indicators using machine vision

Björn Hemming, Heikki Lehto

Research output: Contribution to journalArticleScientificpeer-review

10 Citations (Scopus)

Abstract

Using automatic machine vision-based systems, the calibration of measuring instruments can be extended. With machine vision it is possible to check hundreds of points on the scale of a dial indicator, giving new insight into its sources of error. This paper describes a machine vision-based system for the calibration of dial indicators developed at the Centre for Metrology and Accreditation in Finland, with emphasis on the calculation of measurement uncertainty.
Original languageEnglish
Pages (from-to)45-49
JournalMeasurement Science and Technology
Volume13
Issue number1
DOIs
Publication statusPublished - 2002