Comparison of the radiation temperature scales between MIKES and PTB

  • M. Ojanen*
  • , K. Anhalt
  • , J. Hartmann
  • , S. Schiller
  • , T. Weckström
  • , P. Kärhä
  • , M. Heinonen
  • , E. Ikonen
  • *Corresponding author for this work

    Research output: Contribution to journalArticleScientificpeer-review

    3 Citations (Scopus)

    Abstract

    The radiation temperature scales of PTB and MIKES were compared in the range of 1570-2770 K using four filter radiometers of MIKES, one filter radiometer of PTB, and linear radiation thermometers of both MIKES and PTB. The agreement was partial: two filter radiometers and the linear radiation thermometer of MIKES agreed well with the equipment of PTB, while two filter radiometers deviated from the other equipment. To get a deeper understanding of the reasons of the deviation, the results were studied in terms of both temperature and radiance. The deviation was found constant in terms of radiance. A correction factor was calculated for the radiance level with help of a temperature fixed-point calibration. The use of the correction factor improved the results.

    Original languageEnglish
    Pages (from-to)183-189
    Number of pages7
    JournalMeasurement
    Volume43
    Issue number2
    DOIs
    Publication statusPublished - 2010
    MoE publication typeA1 Journal article-refereed

    Keywords

    • Filter radiometer
    • Intercomparison
    • Linear radiation thermometer
    • Radiation temperature scale