Abstract

An optical imaging system for milliKelvin temperatures has been developed based on a regular B/W surveillance camera (25 frames/s), with its CCD sensor inside the 4-K vacuum can of our nuclear demagnetization cryostat. The heat leak to the nuclear stage, caused by the operation of the video camera, was reduced below 1 nW by careful rf shielding. The construction of the system and its limits of operation are discussed.