Abstract
To investigate, select and publish an LED reference spectrum to complement the CIE Standard Illuminant A in photometric calibrations, a new technical committee has been recently created by CIE. This technical committee requires defining a spectral mismatch index to quantify the match of spectral power distributions (SPDs) of real sources to the selected reference spectrum. A proposal for such an index is given here, assuming that the spectral mismatch index of the SPD of a real source is directly related to the spectral mismatch systematic error introduced when the photometric quantity is measured using a photometer which is calibrated to a light source with the reference SPD. Finally, the value of this index is discussed and interpreted in photometric terms.
Original language | English |
---|---|
Title of host publication | Proceedings of the 29th Session of the CIE |
Publisher | CIE International Commission on Illumination |
Pages | 85-92 |
Volume | 1 |
ISBN (Electronic) | 9783902842749 |
DOIs | |
Publication status | Published - Jun 2019 |
MoE publication type | A4 Article in a conference publication |
Event | 29th Session of the CIE - Washington, United States Duration: 17 Jun 2019 → 19 Jun 2019 |
Conference
Conference | 29th Session of the CIE |
---|---|
Country/Territory | United States |
City | Washington |
Period | 17/06/19 → 19/06/19 |