A low frost-point generator (INRIM 03) operating at sub-atmospheric pressure has been designed and constructed at the Istituto Nazionale di Ricerca Metrologica (INRIM) as part of a calibration facility for upper-air sounding instruments. This new humidity generator covers the frost-point temperature range between −99 °C and −20 °C and works at any controlled pressure between 200 hPa and 1100 hPa, achieving a complete saturation of the carrier gas (nitrogen) in a single passage through a stainless steel isothermal saturator. The generated humid gas contains a water vapour amount fraction between 14 × 10−9 mol mol−1 and 5 × 10−3 mol mol−1. In this work the design of the generator is reported together with characterisation and performance evaluation tests. A preliminary validation of the INRIM 03 against one of the INRIM humidity standards in the common region is also included. Based on experimental test results, an initial uncertainty evaluation of the generated frost-point temperature, Tfp, and water vapour amount fraction, xw, in the limited range down to −75 °C at atmospheric pressure is reported. For the frost-point temperature, the uncertainty budget yields a total expanded uncertainty (k = 2) of less than 0.028 °C, while for the mole fraction the budget yields a total expanded uncertainty of less than 10−6 mol mol−1.
| Original language | English |
|---|---|
| Article number | 054002 |
| Journal | Measurement Science and Technology |
| Volume | 29 |
| Issue number | 5 |
| DOIs | |
| Publication status | Published - 16 Mar 2018 |
| MoE publication type | A1 Journal article-refereed |
Keywords
- frost point
- humidity generator
- hygrometry
- upper-air sensors
Fingerprint
Dive into the research topics of 'Development of a low frost-point generator operating at sub-atmospheric pressure'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver