Abstract
In boron neutron capture therapy (BNCT) the physical dose
must have a metrologically traceable link to the
international dosimetry standards. The dosimetry
procedure at FIR 1 can he considered to guarantee that
the delivered physical doses in the planned treatments of
glioma patients at FIR 1 have a metrologically traceable
links to the dosimetry standards. The dosimetry procedure
is divided into three sections: pre-irradiation methods,
dosimetric monitoring during the irradiation, and the
post-irradiation measurements. The uncertainties in the
physical dose components in the treatments can be
estimated based on the phantom irradiations.
| Original language | English |
|---|---|
| Pages (from-to) | 388-389 |
| Number of pages | 2 |
| Journal | Medical and Biological Engineering and Computing |
| Volume | 37 |
| Issue number | Supplement 1 |
| Publication status | Published - 1999 |
| MoE publication type | Not Eligible |
| Event | 11th Nordic-Baltic Conference on Biomedical Engineering, NBCBM - Tallinn, Estonia Duration: 6 Jun 1999 → 10 Jun 1999 Conference number: 11 |
Fingerprint
Dive into the research topics of 'Dosimetry chain for the treatments of glioma patients in the epithermal neutron beam at the Finnish BNCT facility (FiR 1)'. Together they form a unique fingerprint.Cite this