Dosimetry chain for the treatments of glioma patients in the epithermal neutron beam at the Finnish BNCT facility (FiR 1)

  • Sauli Savolainen
  • , Iiro Auterinen
  • , Carita Aschan
  • , Pekka Hiismäki
  • , Mika Kortesniemi
  • , Antti Kosunen
  • , Petri Kotiluoto
  • , Juha Lampinen
  • , Tiina Seppälä
  • , Tom Seren
  • , Vesa Tanner
  • , Matti Toivonen

Research output: Contribution to journalOther journal contributionScientific

Abstract

In boron neutron capture therapy (BNCT) the physical dose must have a metrologically traceable link to the international dosimetry standards. The dosimetry procedure at FIR 1 can he considered to guarantee that the delivered physical doses in the planned treatments of glioma patients at FIR 1 have a metrologically traceable links to the dosimetry standards. The dosimetry procedure is divided into three sections: pre-irradiation methods, dosimetric monitoring during the irradiation, and the post-irradiation measurements. The uncertainties in the physical dose components in the treatments can be estimated based on the phantom irradiations.