Effects of Soft Encapsulation on the Receive Performance of PMUTs for Implantable Devices

  • Andrada I. Velea
  • , Raphael Panskus
  • , Benedikt Szabo
  • , Vera A. -L. Oppelt
  • , Lukas Holzapfel
  • , Cyril Karuthedath
  • , T.S. Abhilash
  • , Thomas Stieglitz
  • , Alessandro S. Savoia
  • , Vasiliki Giagka

Research output: Contribution to journalArticleScientificpeer-review

2 Citations (Scopus)

Abstract

Ultrasound (US) is a promising modality for wirelessly powering implantable devices, requiring encapsulated receivers to ensure long-term stability. Traditional hermetic packaging often limits acoustic transmission, making polymer-based encapsulation a more suitable alternative. This study investigates how implant-grade polymers, thermoplastic polyurethane, parylene-C, and medical-grade silicones (MED-1000, MED2-4213), affect the receive performance of piezoelectric micromachined ultrasonic transducers (PMUTs). Simulations and measurements between 1 and 7 MHz show that all tested materials exhibit transmission coefficients above 94 % at nm- and μm-scale thicknesses, confirming their acoustic transparency. The results show that although coated PMUTs are acoustically well matched with the surrounding water medium, the added mechanical load of the coating can hinder membrane motion and reduce the energy transferred to the PMUTs. Modeling and experimental data demonstrate that stiffer coatings, such as parylene-C, lead to a reduced sensitivity when similar thicknesses are used. Likewise, residual stress in materials like MED-1000 can also degrade performance. These effects are not evident from acoustic transmission measurements alone, underscoring the need to assess both acoustic and mechanical properties when selecting encapsulation materials. In general, softer materials offer excellent acoustic performance for PMUT encapsulation, while stiffer materials must be applied in thinner layers to avoid impairing PMUT function.
Original languageEnglish
Pages (from-to)1282-1292
Number of pages11
JournalIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control
Volume72
Issue number9
DOIs
Publication statusPublished - 2025
MoE publication typeA1 Journal article-refereed

Funding

This work was supported in part by the Moore4Medical Project through the Electronics Components and Systems for European Leadership (ECSEL) Joint Undertaking under Grant H2020-ECSEL-2019-IA-876190 and in part by the Fraunhofer PREPARE Project DUSTIN.

Keywords

  • Acoustic characterization
  • soft encapsulation
  • piezoelectric micromachined ultrasonic transducer (PMUT)
  • receive performance