Abstract
This is a report to the Technical Committee for Thermometry of EURAMET (EURAMET TC-T) on the regional key comparison EURAMET.T-K9 of SPRT calibration on the International Temperature Scale of 1990 (ITS-90), from 83.8058 K (the Ar triple point) to 692.677 K (the Zn fixed point). The transfer standards used are long-stem SPRTs.
EURAMET.T-K9 is an RMO key comparison linked to the CIPM key comparison CCT-K9. The comparison is organized in five separate collapsed Stars, each one having a CCT-K9 participant as pilot, which provides the link to the KCRV of CCT-K9. Four to five participating laboratories are involved in each Star. The pilots of the five Stars are LNE-Cnam, PTB, INRiM, VSL, NPL. LNE-Cnam is also the coordinator of the comparison, in charge of the collection of the measurement reports from all participants and the preparation of the final report.
EURAMET.T-K9 is an RMO key comparison linked to the CIPM key comparison CCT-K9. The comparison is organized in five separate collapsed Stars, each one having a CCT-K9 participant as pilot, which provides the link to the KCRV of CCT-K9. Four to five participating laboratories are involved in each Star. The pilots of the five Stars are LNE-Cnam, PTB, INRiM, VSL, NPL. LNE-Cnam is also the coordinator of the comparison, in charge of the collection of the measurement reports from all participants and the preparation of the final report.
| Original language | English |
|---|---|
| Article number | 03005 |
| Journal | Metrologia |
| Volume | 61 |
| Issue number | 1A |
| DOIs | |
| Publication status | Published - 14 Oct 2024 |
| MoE publication type | A2 Review article in a scientific journal |