Evaluations of commercial ion selective electrodes: 3. The ammonia gas-sensing membrane probe

Rauno Virtanen

Research output: Book/ReportReport

Abstract

Two makes of ammonia gas sensing membrane probes were evaluated. Calibration curves were determined repeatedly during three months both in separate solutions and in flowing solutions and in the air gap mode.Response times by a step change of analyte concentration were determined.Only minor differences between different makes were found.The measuring range of the probe is about from 1 mol/dm3 to 10 6 mol/dm3.The response time is dependent on concentration level so that at higher concentrations the probes respond considerably faster than at lower concentrations.A marked decrease in response time occurs at the concentration range of 10 4 10 5 mol/dm3.Response times are of few seconds to several minutes.The ammonia probes were found convenient and useful devices though their manipulation requires a certain degree of care.
Original languageEnglish
Place of PublicationEspoo
PublisherVTT Technical Research Centre of Finland
Number of pages22
ISBN (Print)951-38-1336-3
Publication statusPublished - 1981
MoE publication typeNot Eligible

Publication series

SeriesTechnical Research Centre of Finland. Publications
Number6
ISSN0358-5069

Fingerprint

Ion selective electrodes
Ammonia
Gases
Membranes
Calibration
Air

Cite this

Virtanen, R. (1981). Evaluations of commercial ion selective electrodes: 3. The ammonia gas-sensing membrane probe. Espoo: VTT Technical Research Centre of Finland. Technical Research Centre of Finland. Publications , No. 6
Virtanen, Rauno. / Evaluations of commercial ion selective electrodes : 3. The ammonia gas-sensing membrane probe. Espoo : VTT Technical Research Centre of Finland, 1981. 22 p. (Technical Research Centre of Finland. Publications ; No. 6).
@book{71d0fe64c94c4a47b2b455c5baf0f6e5,
title = "Evaluations of commercial ion selective electrodes: 3. The ammonia gas-sensing membrane probe",
abstract = "Two makes of ammonia gas sensing membrane probes were evaluated. Calibration curves were determined repeatedly during three months both in separate solutions and in flowing solutions and in the air gap mode.Response times by a step change of analyte concentration were determined.Only minor differences between different makes were found.The measuring range of the probe is about from 1 mol/dm3 to 10 6 mol/dm3.The response time is dependent on concentration level so that at higher concentrations the probes respond considerably faster than at lower concentrations.A marked decrease in response time occurs at the concentration range of 10 4 10 5 mol/dm3.Response times are of few seconds to several minutes.The ammonia probes were found convenient and useful devices though their manipulation requires a certain degree of care.",
author = "Rauno Virtanen",
year = "1981",
language = "English",
isbn = "951-38-1336-3",
series = "Technical Research Centre of Finland. Publications",
publisher = "VTT Technical Research Centre of Finland",
number = "6",
address = "Finland",

}

Virtanen, R 1981, Evaluations of commercial ion selective electrodes: 3. The ammonia gas-sensing membrane probe. Technical Research Centre of Finland. Publications , no. 6, VTT Technical Research Centre of Finland, Espoo.

Evaluations of commercial ion selective electrodes : 3. The ammonia gas-sensing membrane probe. / Virtanen, Rauno.

Espoo : VTT Technical Research Centre of Finland, 1981. 22 p. (Technical Research Centre of Finland. Publications ; No. 6).

Research output: Book/ReportReport

TY - BOOK

T1 - Evaluations of commercial ion selective electrodes

T2 - 3. The ammonia gas-sensing membrane probe

AU - Virtanen, Rauno

PY - 1981

Y1 - 1981

N2 - Two makes of ammonia gas sensing membrane probes were evaluated. Calibration curves were determined repeatedly during three months both in separate solutions and in flowing solutions and in the air gap mode.Response times by a step change of analyte concentration were determined.Only minor differences between different makes were found.The measuring range of the probe is about from 1 mol/dm3 to 10 6 mol/dm3.The response time is dependent on concentration level so that at higher concentrations the probes respond considerably faster than at lower concentrations.A marked decrease in response time occurs at the concentration range of 10 4 10 5 mol/dm3.Response times are of few seconds to several minutes.The ammonia probes were found convenient and useful devices though their manipulation requires a certain degree of care.

AB - Two makes of ammonia gas sensing membrane probes were evaluated. Calibration curves were determined repeatedly during three months both in separate solutions and in flowing solutions and in the air gap mode.Response times by a step change of analyte concentration were determined.Only minor differences between different makes were found.The measuring range of the probe is about from 1 mol/dm3 to 10 6 mol/dm3.The response time is dependent on concentration level so that at higher concentrations the probes respond considerably faster than at lower concentrations.A marked decrease in response time occurs at the concentration range of 10 4 10 5 mol/dm3.Response times are of few seconds to several minutes.The ammonia probes were found convenient and useful devices though their manipulation requires a certain degree of care.

M3 - Report

SN - 951-38-1336-3

T3 - Technical Research Centre of Finland. Publications

BT - Evaluations of commercial ion selective electrodes

PB - VTT Technical Research Centre of Finland

CY - Espoo

ER -

Virtanen R. Evaluations of commercial ion selective electrodes: 3. The ammonia gas-sensing membrane probe. Espoo: VTT Technical Research Centre of Finland, 1981. 22 p. (Technical Research Centre of Finland. Publications ; No. 6).