Abstract
We introduce an optimized Josephson double array generating two independent voltages up to 1.5V using only one microwave input. Such an array, combined with the quantum Hall resistance, can provide the basic traceability for an electricity metrology laboratory from DC voltage to capacitance. We describe the array and present preliminary experimental results from DC measurements. Extending the output voltage up to 10V is discussed
Original language | English |
---|---|
Title of host publication | 2004 Conference on Precision Electromagnetic Measurements Digest |
Publisher | IEEE Institute of Electrical and Electronic Engineers |
Pages | 154-155 |
ISBN (Electronic) | 978-0-7803-8494-1 |
ISBN (Print) | 978-0-7803-8493-4 |
DOIs | |
Publication status | Published - 2004 |
MoE publication type | A4 Article in a conference publication |
Event | 2004 Conference on Precision Electromagnetic Measurements, CPEM 2004 - London, United Kingdom Duration: 27 Jun 2004 → 2 Jul 2004 |
Conference
Conference | 2004 Conference on Precision Electromagnetic Measurements, CPEM 2004 |
---|---|
Country/Territory | United Kingdom |
City | London |
Period | 27/06/04 → 2/07/04 |