Final report on EURAMET.T-K6.1: Bilateral comparison of the realisations of local dew/frost-point temperature scales in the range -70°C to +20°C

Martti Heinonen, Davor Zvizdic, Danijel Sestan

    Research output: Contribution to journalArticleScientificpeer-review

    Abstract

    As the European extension of the first CCT humidity key comparison, EUROMET.T-K6 was successfully completed in year 2008. After this comparison, a new low dew-point generator was introduced at LPM in Croatia as a result of progress in the EUROMET P912 project. With this new facility, the LPM uncertainties decreased significantly and the operating range became significantly wider. Therefore, it was decided to arrange a bilateral comparison between LPM and MIKES in Finland providing a link to EUROMET.T-K6 and CCT-K6. This comparison was carried out in a manner similar to other K6 comparisons but only one transfer standard was used instead of two units and the measurement point −70 °C was added to the measurement scheme. At all measurement points, the bilateral equivalence was well within the estimated expanded uncertainty at the approximately 95% confidence level. Also, the deviations of the LPM results from the EUROMET.T-K6 reference values were smaller than their expanded uncertainties.
    Original languageEnglish
    Number of pages40
    JournalMetrologia
    Volume50
    Issue number1A
    DOIs
    Publication statusPublished - 2013
    MoE publication typeA1 Journal article-refereed

    Keywords

    • measurement
    • temperatures

    Fingerprint

    Dive into the research topics of 'Final report on EURAMET.T-K6.1: Bilateral comparison of the realisations of local dew/frost-point temperature scales in the range -70°C to +20°C'. Together they form a unique fingerprint.

    Cite this