Frequency comb generator for calibration of laser frequency standards

M. Merimaa*, K. Nyholm, M. Vainio

*Corresponding author for this work

    Research output: Chapter in Book/Report/Conference proceedingConference article in proceedingsScientific

    1 Citation (Scopus)

    Abstract

    We describe the frequency comb generator at MIKES, with which laser frequency standards at 633 nm and 532 nm were measured. Frequencies were found to be within 0.2 kHz from the CIPM values for R(56)32-0 a10 and R(127)11-5 a16. An optical amplifier was developed for measurement of 633 nm standards.