International comparison of He-Ne lasers stabilized with 127I2 at λ ≈ 543 nm (December 1999)

H. R. Simonsen, J. Hu, Kaj Nyholm

    Research output: Contribution to journalArticleScientificpeer-review

    4 Citations (Scopus)

    Abstract

    This second international comparison between He-Ne lasers at 543 nm exhibits similar results to the first international comparison. A standard uncertainty of 4 kHz was obtained against 2 kHz in 1993, while the average frequency difference between the lasers was -1.0 kHz against 14.2 kHz in 1993. The experimental data indicate however that a standard uncertainty below 1 kHz can be expected between two lasers. The frequency difference between the two lasers clearly lies within the limits (2.5 parts in 1011) given in the revised practical realization of the definition of the metre (1997) for the 633 nm He-Ne laser.

    Original languageEnglish
    Pages (from-to)709-714
    JournalMetrologia
    Volume37
    Issue number6
    DOIs
    Publication statusPublished - 1 Dec 2000
    MoE publication typeA1 Journal article-refereed

    Fingerprint

    Dive into the research topics of 'International comparison of He-Ne lasers stabilized with 127I2 at λ ≈ 543 nm (December 1999)'. Together they form a unique fingerprint.

    Cite this