New primary dew-point generators at HMI/FSB-LPM in the range from -70°C to +60°C

Davor Zvizdic (Corresponding Author), Martti Heinonen, Danijel Sestan

Research output: Contribution to journalArticleScientificpeer-review

8 Citations (Scopus)

Abstract

To extend the dew-point range and to improve the uncertainties of the humidity scale realization at HMI/FSB-LPM, new primary low- and high-range dew-point generators were developed and implemented in cooperation with MIKES, in 2009 through EUROMET Project No. 912. The low-range saturator is designed for primary realization of the dew-point temperature scale from −70 °C to + 5 °C, while the high-range saturator covers the range from 1 °C to 60 °C. The system is designed as a single-pressure, single-pass dew-point generator. MIKES designed and constructed both the saturators to be implemented in dew-point calibration systems at LPM. The LPM took care of purchasing and adapting liquid baths, of implementing the temperature and pressure measurement equipment appropriate for use in the systems, and development of gas preparation and flow control systems as well as of the computer-based automated data acquisition. The principle and the design of the generator are described in detail and schematically depicted. The tests were performed at MIKES to investigate how close both the saturators are to an ideal saturator. Results of the tests show that both the saturators are efficient enough for a primary realization of the dew-point temperature scale from −70 °C to + 60 °C, in the specified flow-rate ranges. The estimated standard uncertainties due to the non-ideal saturation efficiency are between 0.02 °C and 0.05 °C.
Original languageEnglish
Pages (from-to)1536-1549
Number of pages13
JournalInternational Journal of Thermophysics
Volume33
Issue number8-9
DOIs
Publication statusPublished - 2012
MoE publication typeA1 Journal article-refereed

Fingerprint

dew point
generators
temperature scales
pressure measurement
data acquisition
temperature measurement
humidity
baths
flow velocity
saturation
preparation
liquids
gases

Keywords

  • dew-point generators
  • humidity
  • saturators

Cite this

@article{971895b93ab24ce5996a34d3ece4c453,
title = "New primary dew-point generators at HMI/FSB-LPM in the range from -70°C to +60°C",
abstract = "To extend the dew-point range and to improve the uncertainties of the humidity scale realization at HMI/FSB-LPM, new primary low- and high-range dew-point generators were developed and implemented in cooperation with MIKES, in 2009 through EUROMET Project No. 912. The low-range saturator is designed for primary realization of the dew-point temperature scale from −70 °C to + 5 °C, while the high-range saturator covers the range from 1 °C to 60 °C. The system is designed as a single-pressure, single-pass dew-point generator. MIKES designed and constructed both the saturators to be implemented in dew-point calibration systems at LPM. The LPM took care of purchasing and adapting liquid baths, of implementing the temperature and pressure measurement equipment appropriate for use in the systems, and development of gas preparation and flow control systems as well as of the computer-based automated data acquisition. The principle and the design of the generator are described in detail and schematically depicted. The tests were performed at MIKES to investigate how close both the saturators are to an ideal saturator. Results of the tests show that both the saturators are efficient enough for a primary realization of the dew-point temperature scale from −70 °C to + 60 °C, in the specified flow-rate ranges. The estimated standard uncertainties due to the non-ideal saturation efficiency are between 0.02 °C and 0.05 °C.",
keywords = "dew-point generators, humidity, saturators",
author = "Davor Zvizdic and Martti Heinonen and Danijel Sestan",
year = "2012",
doi = "10.1007/s10765-012-1211-3",
language = "English",
volume = "33",
pages = "1536--1549",
journal = "International Journal of Thermophysics",
issn = "0195-928X",
publisher = "Springer",
number = "8-9",

}

New primary dew-point generators at HMI/FSB-LPM in the range from -70°C to +60°C. / Zvizdic, Davor (Corresponding Author); Heinonen, Martti; Sestan, Danijel.

In: International Journal of Thermophysics, Vol. 33, No. 8-9, 2012, p. 1536-1549.

Research output: Contribution to journalArticleScientificpeer-review

TY - JOUR

T1 - New primary dew-point generators at HMI/FSB-LPM in the range from -70°C to +60°C

AU - Zvizdic, Davor

AU - Heinonen, Martti

AU - Sestan, Danijel

PY - 2012

Y1 - 2012

N2 - To extend the dew-point range and to improve the uncertainties of the humidity scale realization at HMI/FSB-LPM, new primary low- and high-range dew-point generators were developed and implemented in cooperation with MIKES, in 2009 through EUROMET Project No. 912. The low-range saturator is designed for primary realization of the dew-point temperature scale from −70 °C to + 5 °C, while the high-range saturator covers the range from 1 °C to 60 °C. The system is designed as a single-pressure, single-pass dew-point generator. MIKES designed and constructed both the saturators to be implemented in dew-point calibration systems at LPM. The LPM took care of purchasing and adapting liquid baths, of implementing the temperature and pressure measurement equipment appropriate for use in the systems, and development of gas preparation and flow control systems as well as of the computer-based automated data acquisition. The principle and the design of the generator are described in detail and schematically depicted. The tests were performed at MIKES to investigate how close both the saturators are to an ideal saturator. Results of the tests show that both the saturators are efficient enough for a primary realization of the dew-point temperature scale from −70 °C to + 60 °C, in the specified flow-rate ranges. The estimated standard uncertainties due to the non-ideal saturation efficiency are between 0.02 °C and 0.05 °C.

AB - To extend the dew-point range and to improve the uncertainties of the humidity scale realization at HMI/FSB-LPM, new primary low- and high-range dew-point generators were developed and implemented in cooperation with MIKES, in 2009 through EUROMET Project No. 912. The low-range saturator is designed for primary realization of the dew-point temperature scale from −70 °C to + 5 °C, while the high-range saturator covers the range from 1 °C to 60 °C. The system is designed as a single-pressure, single-pass dew-point generator. MIKES designed and constructed both the saturators to be implemented in dew-point calibration systems at LPM. The LPM took care of purchasing and adapting liquid baths, of implementing the temperature and pressure measurement equipment appropriate for use in the systems, and development of gas preparation and flow control systems as well as of the computer-based automated data acquisition. The principle and the design of the generator are described in detail and schematically depicted. The tests were performed at MIKES to investigate how close both the saturators are to an ideal saturator. Results of the tests show that both the saturators are efficient enough for a primary realization of the dew-point temperature scale from −70 °C to + 60 °C, in the specified flow-rate ranges. The estimated standard uncertainties due to the non-ideal saturation efficiency are between 0.02 °C and 0.05 °C.

KW - dew-point generators

KW - humidity

KW - saturators

U2 - 10.1007/s10765-012-1211-3

DO - 10.1007/s10765-012-1211-3

M3 - Article

VL - 33

SP - 1536

EP - 1549

JO - International Journal of Thermophysics

JF - International Journal of Thermophysics

SN - 0195-928X

IS - 8-9

ER -