New primary low-range dew-point generator at LPM

Davor Zvizdic, Martti Heinonen, Tomislav Veliki, Daniel Sestan

Research output: Chapter in Book/Report/Conference proceedingConference article in proceedingsScientificpeer-review

5 Citations (Scopus)

Abstract

This paper describes the new primary low-range dew-point generator at LPM developed in cooperation with MIKES through EUROMET project no. 912. The generator is designed for primary realisation of the dew-point temperature scale from -70 °C to +5 °C. The principle and the design of the generator are described in detail and schematically depicted. Results of the preliminary efficiency tests of it's saturator show that it is efficient enough for a primary realisation of the dew-point temperature scale. Standard uncertainties due to the non-ideal saturation efficiency are estimated to be 0.02 °C.

Original languageEnglish
Title of host publication19th IMEKO World Congress 2009
Pages2011-2014
Number of pages4
Volume4
Publication statusPublished - 2009
MoE publication typeA4 Article in a conference publication
Event19th IMEKO World Congress 2009 - Lisbon, Portugal
Duration: 6 Sep 200911 Sep 2009

Conference

Conference19th IMEKO World Congress 2009
CountryPortugal
CityLisbon
Period6/09/0911/09/09

Keywords

  • Dew-point generator
  • Humidity
  • Saturator

Fingerprint Dive into the research topics of 'New primary low-range dew-point generator at LPM'. Together they form a unique fingerprint.

  • Cite this

    Zvizdic, D., Heinonen, M., Veliki, T., & Sestan, D. (2009). New primary low-range dew-point generator at LPM. In 19th IMEKO World Congress 2009 (Vol. 4, pp. 2011-2014)