New primary low-range dew-point generator at LPM

Davor Zvizdic, Martti Heinonen, Tomislav Veliki, Daniel Sestan

    Research output: Chapter in Book/Report/Conference proceedingConference article in proceedingsScientificpeer-review

    8 Citations (Scopus)

    Abstract

    This paper describes the new primary low-range dew-point generator at LPM developed in cooperation with MIKES through EUROMET project no. 912. The generator is designed for primary realisation of the dew-point temperature scale from -70 °C to +5 °C. The principle and the design of the generator are described in detail and schematically depicted. Results of the preliminary efficiency tests of it's saturator show that it is efficient enough for a primary realisation of the dew-point temperature scale. Standard uncertainties due to the non-ideal saturation efficiency are estimated to be 0.02 °C.

    Original languageEnglish
    Title of host publication19th IMEKO World Congress 2009
    Pages2011-2014
    Number of pages4
    Volume4
    Publication statusPublished - 2009
    MoE publication typeA4 Article in a conference publication
    Event19th IMEKO World Congress 2009 - Lisbon, Portugal
    Duration: 6 Sept 200911 Sept 2009

    Conference

    Conference19th IMEKO World Congress 2009
    Country/TerritoryPortugal
    CityLisbon
    Period6/09/0911/09/09

    Keywords

    • Dew-point generator
    • Humidity
    • Saturator

    Fingerprint

    Dive into the research topics of 'New primary low-range dew-point generator at LPM'. Together they form a unique fingerprint.