Abstract
We demonstrate an off-axis cavity ring-down spectroscopy system, which uses a mid-infrared continuous- wave optical parametric oscillator as a light source. Off-axis injectionwith re-entrant configuration of the ring-down cav- ity is used to achieve high spectral resolution while main- taining high measurement speed. This makes the setup suit- able for sensitive molecular spectroscopy in the mid-infrared region, particularly for studies that require high temporal resolution. Formaldehyde (H2CO) absorption spectrum at 3.4 µmismeasured using the off-axis re-entrant cavity ring- down spectrometer.
| Original language | English |
|---|---|
| Pages (from-to) | 839-847 |
| Journal | Applied Physics B: Lasers and Optics |
| Volume | 107 |
| Issue number | 3 |
| DOIs | |
| Publication status | Published - 25 May 2012 |
| MoE publication type | A1 Journal article-refereed |
Fingerprint
Dive into the research topics of 'Off-axis re-entrant cavity ring-down spectroscopy with a mid-infrared continuous-wave optical parametric oscillator'. Together they form a unique fingerprint.Cite this