Quality assurance procedures for the neutron beam monitors at the FiR 1 BNCT facility

Iiro Auterinen, Tom Serén, Petri Kotiluoto, Jouni Uusi-Simola, Sauli Savolainen

Research output: Contribution to journalArticleScientificpeer-review

5 Citations (Scopus)

Abstract

In order to assure the stability of the beam, the reliability of the beam monitoring system and the quality of the patient dose delivered, several procedures are followed at the FiR 1 epithermal beam in Finland. Routine procedures include in-phantom activation measurements before each patient treatment and a long-term follow-up of the results. The sensitivity of the beam monitors to external objects in the beam and to variations in the control rod positions in the reactor has been checked and found insignificant. The linearity of the beam monitor channels has been checked with activation measurements. It was found that due to saturation effects a correction of 11% has to be applied when extrapolating results from experiments at low power to full power using the reference monitor channel. The correction is even larger for other channels with higher count rates.

Original languageEnglish
Pages (from-to)1015-1019
Number of pages5
JournalApplied Radiation and Isotopes
Volume61
Issue number5
DOIs
Publication statusPublished - 1 Nov 2004
MoE publication typeA1 Journal article-refereed

Fingerprint

neutron beams
assurance
monitors
control rods
activation
Finland
linearity
reactors
saturation
dosage
sensitivity

Keywords

  • Beam monitoring
  • BNCT
  • Non-linearity
  • Quality assurance

Cite this

Auterinen, Iiro ; Serén, Tom ; Kotiluoto, Petri ; Uusi-Simola, Jouni ; Savolainen, Sauli. / Quality assurance procedures for the neutron beam monitors at the FiR 1 BNCT facility. In: Applied Radiation and Isotopes. 2004 ; Vol. 61, No. 5. pp. 1015-1019.
@article{26fd633116094bdd9deb00b3583e7d82,
title = "Quality assurance procedures for the neutron beam monitors at the FiR 1 BNCT facility",
abstract = "In order to assure the stability of the beam, the reliability of the beam monitoring system and the quality of the patient dose delivered, several procedures are followed at the FiR 1 epithermal beam in Finland. Routine procedures include in-phantom activation measurements before each patient treatment and a long-term follow-up of the results. The sensitivity of the beam monitors to external objects in the beam and to variations in the control rod positions in the reactor has been checked and found insignificant. The linearity of the beam monitor channels has been checked with activation measurements. It was found that due to saturation effects a correction of 11{\%} has to be applied when extrapolating results from experiments at low power to full power using the reference monitor channel. The correction is even larger for other channels with higher count rates.",
keywords = "Beam monitoring, BNCT, Non-linearity, Quality assurance",
author = "Iiro Auterinen and Tom Ser{\'e}n and Petri Kotiluoto and Jouni Uusi-Simola and Sauli Savolainen",
year = "2004",
month = "11",
day = "1",
doi = "10.1016/j.apradiso.2004.05.034",
language = "English",
volume = "61",
pages = "1015--1019",
journal = "Applied Radiation and Isotopes",
issn = "0969-8043",
publisher = "Elsevier",
number = "5",

}

Quality assurance procedures for the neutron beam monitors at the FiR 1 BNCT facility. / Auterinen, Iiro; Serén, Tom; Kotiluoto, Petri; Uusi-Simola, Jouni; Savolainen, Sauli.

In: Applied Radiation and Isotopes, Vol. 61, No. 5, 01.11.2004, p. 1015-1019.

Research output: Contribution to journalArticleScientificpeer-review

TY - JOUR

T1 - Quality assurance procedures for the neutron beam monitors at the FiR 1 BNCT facility

AU - Auterinen, Iiro

AU - Serén, Tom

AU - Kotiluoto, Petri

AU - Uusi-Simola, Jouni

AU - Savolainen, Sauli

PY - 2004/11/1

Y1 - 2004/11/1

N2 - In order to assure the stability of the beam, the reliability of the beam monitoring system and the quality of the patient dose delivered, several procedures are followed at the FiR 1 epithermal beam in Finland. Routine procedures include in-phantom activation measurements before each patient treatment and a long-term follow-up of the results. The sensitivity of the beam monitors to external objects in the beam and to variations in the control rod positions in the reactor has been checked and found insignificant. The linearity of the beam monitor channels has been checked with activation measurements. It was found that due to saturation effects a correction of 11% has to be applied when extrapolating results from experiments at low power to full power using the reference monitor channel. The correction is even larger for other channels with higher count rates.

AB - In order to assure the stability of the beam, the reliability of the beam monitoring system and the quality of the patient dose delivered, several procedures are followed at the FiR 1 epithermal beam in Finland. Routine procedures include in-phantom activation measurements before each patient treatment and a long-term follow-up of the results. The sensitivity of the beam monitors to external objects in the beam and to variations in the control rod positions in the reactor has been checked and found insignificant. The linearity of the beam monitor channels has been checked with activation measurements. It was found that due to saturation effects a correction of 11% has to be applied when extrapolating results from experiments at low power to full power using the reference monitor channel. The correction is even larger for other channels with higher count rates.

KW - Beam monitoring

KW - BNCT

KW - Non-linearity

KW - Quality assurance

UR - http://www.scopus.com/inward/record.url?scp=4143051428&partnerID=8YFLogxK

U2 - 10.1016/j.apradiso.2004.05.034

DO - 10.1016/j.apradiso.2004.05.034

M3 - Article

C2 - 15308185

AN - SCOPUS:4143051428

VL - 61

SP - 1015

EP - 1019

JO - Applied Radiation and Isotopes

JF - Applied Radiation and Isotopes

SN - 0969-8043

IS - 5

ER -