Results on the effective area of a DHI piston-cylinder unit with the nominal area of 196 mm²

Markku Rantanen, Sari Semenoja

Research output: Book/ReportReport

Abstract

The pressure laboratory of the Centre for Metrology and Accreditation (MIKES) purchased in 2004 a new piston cylinder unit with the nominal effective area of 196 mm2 to be used as a working standard in absolute and gauge pressure ranges from 25 kPa to 1,75 MPa. A series of measurements on the effective area of the new unit was carried on in MIKES, in Swedish National Testing and Research Institute (SP) and in Metrosert, the leading pressure laboratory in Estonia. The results from MIKES, SP and Metrosert were in a good agreement with each other and with the value given by the manufacturer, DH Instruments, Inc., USA. There seems to be no significant difference in effective areas obtained in gauge and absolute modes.
Original languageEnglish
Place of PublicationHelsinki
PublisherCentre of Metrology and Accreditation (MIKES)
Number of pages14
ISBN (Print)952-5209-98-9
Publication statusPublished - 2005
MoE publication typeD4 Published development or research report or study

Publication series

SeriesMIKES Publication
NumberJ5/2005
ISSN1235-5704

Keywords

  • metrology
  • pressure calibration

Fingerprint

Dive into the research topics of 'Results on the effective area of a DHI piston-cylinder unit with the nominal area of 196 mm²'. Together they form a unique fingerprint.