Abstract
Analytic expressions for the ratio uncertainty of the overlapped-tube cryogenic current comparator are derived for dimensioning the overlapped-tube comparator. Theoretical predictions agree reasonably well with experimental observations by H. Ramin (Arch. Elektrotech., vol.58, p.53, 1976). The construction of a very sensitive cryogenic current comparator and the possibility of minimizing the ratio uncertainty by reducing the size of the detection coil are also discussed.<>
Original language | English |
---|---|
Pages (from-to) | 689-697 |
Journal | IEEE Transactions on Instrumentation and Measurement |
Volume | 39 |
Issue number | 5 |
DOIs | |
Publication status | Published - 1990 |