Abstract
Analytic expressions for the ratio uncertainty of the overlapped-tube cryogenic current comparator are derived for dimensioning the overlapped-tube comparator. Theoretical predictions agree reasonably well with experimental observations by H. Ramin (Arch. Elektrotech., vol.58, p.53, 1976). The construction of a very sensitive cryogenic current comparator and the possibility of minimizing the ratio uncertainty by reducing the size of the detection coil are also discussed.<>
| Original language | English |
|---|---|
| Pages (from-to) | 689-697 |
| Journal | IEEE Transactions on Instrumentation and Measurement |
| Volume | 39 |
| Issue number | 5 |
| DOIs | |
| Publication status | Published - 1990 |
| MoE publication type | A1 Journal article-refereed |
Fingerprint
Dive into the research topics of 'The ratio error of the overlapped-tube cryegenic current comparator'. Together they form a unique fingerprint.Cite this