Abstract
A prototype of SQUID read-out for current measurements at the output of a normal metal hot-electron microbolometer has been developed and tested. The system is based on serial VTT dc SQUID and input μ-metal core solenoid transformer. The achieved current resolution is 300 fA/Hz1/2. Johnson noise of metal resistor and shot noise of tunnel junction were used for current calibration of SQUID read-out. The current noise spectra of 35 k SIN tunnel junction measured at different bias voltages are presented.
| Original language | English |
|---|---|
| Pages (from-to) | 161-165 |
| Journal | Physica C: Superconductivity and its Applications |
| Volume | 368 |
| Issue number | 1-4 |
| DOIs | |
| Publication status | Published - 2002 |
| MoE publication type | A1 Journal article-refereed |
Keywords
- dc SQUID
- SIN tunnel junction
- SQUID femtoamperemeter