Abstract
We addressed the question of how to assign confidence limits to the measured trajectory of a moving MEMS membrane. This was done by calibrating our SSWLI setup using a traceable piezoelectric transducer as a transfer standard. We calibrated our SSWLI setup for quasi-dynamic measurements of vertical movements up to 14 μm with standard uncertainty of 9.4 nm.
Original language | English |
---|---|
Title of host publication | Fringe 2013 |
Subtitle of host publication | 7th International Workshop on Advanced Optical Imaging and Metrology |
Publisher | Springer |
Pages | 491-496 |
ISBN (Electronic) | 978-3-642-36359-7 |
ISBN (Print) | 978-3-642-36358-0 |
DOIs | |
Publication status | Published - 1 Jan 2014 |
MoE publication type | A4 Article in a conference publication |
Event | 7th International Workshop on Advanced Optical Imaging and Metrology, Fringe 2013 - Nurtingen, Germany Duration: 1 Jan 2014 → 1 Jan 2014 |
Conference
Conference | 7th International Workshop on Advanced Optical Imaging and Metrology, Fringe 2013 |
---|---|
Country/Territory | Germany |
City | Nurtingen |
Period | 1/01/14 → 1/01/14 |